SALES 08 8357 8022 | sales@iqmedical.com.au

Megatron S4 HPS specifications

Technical specifications
Dimensions (L x W x H)420 x 420 x 370 mm
Weight25 kg
Power SupplyVoltage 100 – 240 V
Frequency50 / 60 Hz
Power ConsumptionMax. 100-400 V/W ~ 240-400 V/W
Peristaltic Pump
VacuumLinear, exponential and fixed control 1 – 600 mm Hg
AspirationLinear and fixed control 1 – 50 ml / min
Pulsation FrequencyAdjustable from to 10 Hz
Rise TimeAdjustable from 0.5 to 5 sec
Direct Control Pump
VacuumLinear, exponential and fixed control 1 – 600 mm Hg
Rise TimeAdjustable from 0.5 to 5 sec
Cold LightOcuLED connector
Optional Xenon light source (Xenetron III)
Ultrasound
Frequency Range27 – 55 kHz
Power OutputLinear, exponential and fixed control 0 – 100 %
PulsationAdjustable 1 – 60 Hz
Pulse TimeAdjustable 2 – 900 ms
Vitrectomy
Cut RateLinear, exponential and fixed control
Magnetic10 – 2,000 / 4,000* 1 / min
Pneumatic UNO Colorline Mach220 – 12,000 1 / min
Pneumatic UNO Colorline10 – 6,000 1 / min
Air / Oil injection
Air InjectionAdjustable 0 – 100 mm hg
Oil InjectionAdjustable 0.5 – 6 bar
Bipolar DiathermyBipolar Wet Field Coagulation
Endo-diathermy
Power OutputLinear and fixed control 0 – 15 W
PulsationAdjustable 1 – 20 Hz

* with MegaVit G-30910 and G-30985