SALES 08 8357 8022 | sales@iqmedical.com.au

PreciSAL Monofocal specifications

 

IQ Medical_MBI PreciSAL Monofocal

IOL DesignAspheric single piece for posterior capsular bag
MaterialHydrophobic Acrylic with UV absorbers.
Aspheric: Models 302AC and 302A (with violet filter)
UV Cut Off at 10% T> 385 nm (model 302AC), > 392 nm (model 302A)
Material Water Content<0.5%
Refractive Index1.5
ABBE Number50
Q Value-0.11
Optic DesignBiconvex, square edged optic and haptic
Optic Diameter6.0mm
Overall Length13.0 mm
Haptic Angle0°
Dioptre RangeAvailable in powers from 1.0D to 30.0D:
18.0D to 28.0D in 0.25D increments
10.0D to 17.5D & 28.5D to 30.0D in 0.5D increments
1.0D to 9.0D in 1.0D increments
Manufacturing Tolerance +/- 0.125D
Preloaded Mechanism2.2mm Incision (P302AC; P302A
Manufacturer’s A-Constant118.7 (for contact AND immersion biometry)
Recommended A-Constant*SRK II: 119.2
SRK-T: 118.9
sf: 1.75 [Holladay I]
HAIGIS: [a0: 1.32, a1: 0.40, a2: 0.10]
ACD*5.337 [Holladay II]; 5.51 [Hoffer Q]; 5.16 [Manufacturer]
Method of SterilisationEthylene Oxide (ETO)
Rebate CodesIQ013 (302AC); IQ014 (P302AC); IQ015 (302A); IQ016 (P302A)

*These values are shown as guidelines only for use with optical biometry for calculation of implant power. IQ Medical and MBI recommend that surgeons develop their own values based on individual technique, measuring equipment and desired post-operative results. In no way are these values meant to be definitive.